 +~~ODT~~
 +====== Align Sample ======
 +===== Horizontal =====
 +Analog to the other scans, align in the middle of the preview window
 +
 +===== Vertical =====
 +  * Top of the Sample Holder to the Bottom of the Camera window (Note Position e.g. as '​-7850'​)
 +  * Top of the Sample to the Top of the Camera Window, so you have the full sample inside X Camera Windows (Note Position e.g. as '​150'​)
 +
 +====== Stacked scan ======
 +  * Set Name, Angle (-180:180) and Number of Projections in the ControlPanel
 +  * Use the Terminal of '​x02da-cons-2'​ to start the stacked scan with the following command:
 +  ​
 +  /​work/​sls/​bin/​stacked_scan.py -t -7850 150 2048 10 1
 +
 +  * Parameterlist:​
 +    * '​-t'​ just tests all Parameters and doesn'​t run anything. Use this to double-check the positions in the preview
 +    * Pos1 (-7850) and Pos2 (150) are the Y-Positions of the Sample at the top and bottom, in MicroMeters
 +    * '​2048'​ Number of Lines in Projection (in the camera window). A small overlap is built into the script, so you can use the '​real'​ number
 +    * '​10'​ Magnification
 +    * '​1' ​ Binning